STP-200/300/400 Series Instruction Manual
10.3 How to Charge the External Battery
The STP control unit is equipped with a constant voltage (27.3 V) circuit and a
constant current (0.3 A) circuit for charging the battery.
The battery is charged while the power is "ON," regardless of the operation state
of the STP pump.
After a power failure, charge the battery for approx. 12 times the BATTERY
BACKUP time. For example, if the BATTERY BACKUP time is 15 minutes, it
takes approx. 3 hours to charge the battery.
If a power failure occurs 2 times within 3 hours, charge the battery for more than
3 hours before starting the STP pump.
10.4 How to Dispose of the External Battery
Dispose of or collect worn-out batteries in accordance with the guidelines given
by each national and/or local government. If you procure the batteries, dispose of
them according to the guidelines given by the battery manufacturer.
Pay attention to short-circuiting or leakage of the batteries when
disposing of them.
To prevent the batteries from short-circuiting, remove the lead wires
and connectors from the batteries.
Cut the lead wires one by one so as not to cause a short-circuiting.
Coat the conductive part of the batteries with tape.
It is recommended to dispose of worn-out batteries after discharging
them thoroughly.
Batteries manufactured by Japan Storage Battery Co., Ltd. will be
collected by Japan Storage Battery Co., Ltd. For detailed
information, contact Japan Storage Battery Co., Ltd. Or Service
office.