STP-200/300/400 Series Instruction Manual
10.1 Specifications for the External Battery
When you use external batteries, select one in accordance with the specifications
given in Table 10.1.
If you use external batteries which do not conform to the
specifications, battery backup will not function during a power failure
which could result in damage to the STP pump.
Table 10.1 Specifications for the External Battery
Item Specifications
Capacity (Ah) 4.5 or more
Voltage (V) 24
Current (A) 3
Operation time (minutes) Approx. 15
Operation frequency (times) Continues 2 or more
Charging voltage (V) 27.3
It is recommended to use PORTALAC Series Small-Size Sealed Lead-Acid
Battery manufactured by Japan Storage Battery Co., Ltd. as an external battery.