STP-200/300/400 Series Instruction Manual
4.2.2 Installation Area
Leave enough space for the following in addition to that for the STP control unit
(See Figure 4.4, "Peripheral Space of the STP Control Unit.")
• Space for maintenance and inspection
• Space for inlet and outlet of air for cooling
- Top and side: 5 cm or more
- Bottom: 1.8 cm or more (height of the rubber foot)
• Space for connecting the cables
- Rear: 15 cm or more
◇ The minimum bending radius of the STP connection cable is
150 mm (See Figure 16.3, "External Appearance of the STP Control
Unit" [Cable Space]).
DO NOT excessively bend the cables and beware of any obstacles
when installing the STP pump.
Also, leave enough space to install other cables without bending
them excessively.
5cm or more
Rack Wall Side
5cm or more
5cm or more
15cm or more
5cm or more
1.8cm or more
Rack Wall Side
Figure 4.4 Peripheral Space of the STP Control Unit
4-8