STP-200/300/400 Series Instruction Manual
6) Depending upon the type of the backing pump used, oil vapor may
contaminate the inside of the STP pump. Some oil viscosity could cause a
malfunction when there is a strong reverse flow of oil.
Take the following measures to ensure the correct flow of oil:
Attach a vacuum valve to the middle of the piping between the STP pump
outlet port flange and the backing pump.
Attach an absorption trap adjacent to the vacuum valve.
Piping at the Inlet Port Flange
Attach the inlet port to the high vacuum side.
Maximum working pressure:
Pressure at the inlet port flange
applicable continuously
1.3×10
-2
Pa [1×10
-4
Torr]
(for natural air cooled)
Piping at the Outlet Port Flange
Attach the outlet port to the inlet port flange of the backing pump
(primary side pump).
Allowable backing pressure:
Pressure at the outlet port flange
applicable continuously
13
Pa [0.1
Torr]
(for natural air cooled)
To attain the ultimate pressure shown in Table 16.1,”Specifications
for the STP Pump,” set the pressure at the outlet port flange to
1.3 Pa (10
-2
Torr) or less.
3-14