Model 5014i to 5030i Upgrade Procedures
SHARP Connection Bracket Assembly Installation
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-3
1
3
4
2
STEPS
1 = UNFASTEN THUMB NUT
2 = SLIDE TAPE PICKUP SPOOL OFF
3 = SLIDE TAPE ASSEMBLY OUT TO GAIN ACCESS TO SCREW AREA (4)
4 = FASTEN 2X SCREWS TO CONNECTION BRACKET THRU WALL PLATE
5 = MAKE SURE CONNECTOR ORIENTATION CLOSER TO THIS EDGE
See Note 5
Figure 10–2. Tape Pick-up Spool Removal
3. Remove the two Philips screws from the connection bracket and align
the bracket onto the wall plate, as shown above in Step 4, and insert the
two screws through the wall plate and tighten.
4. Reinstall the tape pickup spool in the reverse order of removal in Step 2
by sliding all the spacers, O-rings and rear support guide plate into its
original assembly order.
5. Install the filter tape following the filter tape path shown in Figure 10–
3. Use adhesive tape to adhere the filter tape to the take up spool and
wind the tape over itself at least once.