Servicing
Transformer Replacement
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-37
4. Unplug the RTD connector from the measurement interface board
“flowtemp1”.
5. Carefully install the new RTD by following the previous steps in
reverse. Match the marked black line on the RTD to be flush with the
fitting.
Note The black line is the pre-measured depth for the RTD.
6. Calibrate the instrument. Refer to the “Calibration” chapter in this
manual.
Figure 7–19. Replacing the RTD
Use the following procedure to replace the transformer (Figure 7–20).
Equipment Required:
Transformer assembly
Philips screwdriver, #2
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
Transformer
Replacement
Unplug RTD Connector
Udjustable Wrench