Operation
Service Menu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-93
THERMAL MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0023.000 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Vacuum Mass coefficient screen allows the user to view and set the
vacuum coefficient using manual calibration mode. (Factory set.)
In the Main Menu, choose Service > Mass Calibration > Vacuum
Coefficient.
VACUUM MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0002.000 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Barometer Mass Coefficient screen allows the user to view and set the
barometer coefficient using manual calibration mode. (Factory set.)
In the Main Menu, choose Service > Mass Calibration > Baro
Coefficient.
BARO. MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0.000540 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Mass Coefficient submenu allows the user to view and set the mass
coefficient using manual or auto mode. (Factory set.)
In the Main Menu, choose Service > Mass Calibration > Mass
Coefficient.
Vacuum Mass Coefficient
Barometer Mass
Coefficient
Mass Coefficient