Servicing
External Zeroing Filter
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-53
4. Replace the case cable assembly by following the previous steps in
reverse.
Figure 7–31. Replacing the Case Cable
Use the following procedure to replace the external zeroing filter (Figure 7–
32).
Equipment Required:
Adjustable wrench
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Note the orientation of the arrow on the filter assembly. See Figure 2–7
for orientation.
2. Compress the collar of the quick connect fitting and remove the tubing
from fittings.
3. Install filter assembly by following the previous steps in reverse.