Operation
Alarms Menu
3-80 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
CONC ALARMS:
>AVG PM OK
AVG SHARP OK
AVG NEPH OK
INST PM OK
INST SHARP OK
RANGE AVG DIAGS ALARM
INST NEPH OK
The Average PM screen displays the current average PM concentration and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 0 to 10000 μg/m
3
or 0 to 10 mg/m
3
. If the average PM
concentration goes beyond either the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on the
Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Conc Alarms > Avg PM.
AVERAGE PM CONC:
ACTUAL 0.0
>MIN 0.0
MAX 10000
RANGE AVG DIAGS ALARM
The Minimum Average PM Concentration alarm limit screen is used to
change the minimum average PM concentration alarm limit. The
minimum and maximum average PM concentration screens function the
same way.
In the Main Menu, choose Alarms > Conc Alarms > Avg PM > Min
or Max.
AVERAGE PM CONC:
ACT MIN: 0.0
SET MIN: xxxxx100 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Average PM
Min and Max Average
PM Concentration Limits