Servicing
LCD Module Replacement
7-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Figure 7–23. Replacing the Front Panel Board and the LCD Module
Use the following procedure to replace the LCD module (Figure 7–23).
Equipment Required:
LCD module
Philips screwdriver
CAUTION If the LCD panel breaks, do not let the liquid crystal contact
your skin or clothes. If the liquid crystal contacts your skin or clothes, wash
it off immediately using soap and water.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
Do not remove the LCD panel or frame from the LCD module.
The LCD polarizing plate is very fragile, handle it carefully.
LCD Module
Replacement
Mounting Studs (2)
LCD Module
Retaining Screw
LCD Module Screws (2)
Front Panel Board
Two-Wire Connector
Ribbon Cables (3)
Bottom Support