vapour and CO
2
present at high concentrations in the air. Filters are used to
minimize the effect but humidity should be measured simultaneously to
some tracers, suc h as N
2
O, to allow for corrections.
.
Analysis time – 10–50 s.
.
Accuracy – 1 per cent of full scale if the zero drift is controlled.
Photo-acoustic detector
This analyser is also an infrared absorption spectrometer, but uses a different
detector. An infrared radiation beam is first choppe d then optically filtered to
leave only frequencies that are absorbed by the tracer of interest. This beam
then enters a gas-tight chamber containing the air sample. As above, the
sample is heated and cooled in phase with the chopping frequency, creating
sound waves in the chamber. This is the photo-acoustic effect. Microphones
detect these sound waves.
Characteristics of analysers with photo-acoustic detector:
.
Analysable tracers – N
2
O, SF6, CO
2
, freons F11, F112, 113 and 114, halons
(one filter for each tracer). Other detectable gases are not suitable as tr acers,
such as H
2
O, benzene, alkenes and so on.
.
Sensitivity – the detection limit depends on the tracer but is typically
0.05 ppm, and the dynamic range is 10
5
. The lowest full-scale range may
then be 2 ppm but 10 ppm is recommended with usual tracers. The sensi-
tivity for N
2
O, CO and CO
2
drops strongly when these gases are di luted
in dry nitrogen, as is often the case for calibration gases. Adding a special
‘Nafion’ tube in the sampling circuit allows for the moistening of the mixture
in or der to recover the normal high sensitivity.
.
Interfering gases – several gases (which are not necessarily present in the air)
may interfere with each tracer. Therefore, filters and tracers should be
chosen in accordance with manufacturer’s specifications.
.
Analysis time – 30 s for one gas, 105 s for five gases and H
2
O.
.
Accuracy – 1 per cent of full scale.
Mass spectrometry
The pressure of the air sample is first lowered to about 10
5
Pa by pumping it
through a capillary tube. The molecules of the sample are then ionized, accel-
erated to a given velocity and passed into a mass spectrometer.
The classical mass spectrometer curves the trajectory of the ions with a
strong magnetic field. The radi us of curvature depends on the velocity and
charge-to-mass ratio of the ion; only those having the appro priate combination
will pass a slit placed in front of the detector.
The most suitable spectrometer is, however, the quadrupole mass spectr o-
meter, which is currently used in vacuum processes to analyse the residual
gases. The gases entering into the analyser are ionized and the positive ions
are separated by directing them axially between two pairs of rods creating an
electric field at variable radi o frequency. The ions follow a helicoidal path in
144 Ventilation and Airflow in Buildings
Comentários a estes Manuais