EXT Split Flow Compound Turbomolecular Pumps
TECHNICAL DATA
PAGE
3
Mar 07 3 Issue F
2 TECHNICAL DATA
2.1 General
2.2 Pumping media
Note: Concentrations of gases may be modified by the compression of the pump.
The pumps are designed to pump the following residual gases normally used in high-vacuum systems:
Performance See Table 4
Dimensions See Figures 1, 2, 3 and 4
Maximum inlet-flange temperature 70
o
C
Maximum magnetic field 3.5 mT Horizontal field
Installation category EN61010 part 1, Category 1
Pollution degree EN61010 part 1, Category 2
Table 1 - General
WARNING
Vent dangerous gases and gas mixtures safely. Do not expose people to these gases.
WARNING
Do not use Split Flow Pumps to pump explosive gas mixtures as the pumps are not
suitable for this purpose.
WARNING
If pyrophoric gases are pumped then the customer must supply adequate dilution with an
inert gas.
WARNING
If pumping hazardous gases or vapours, the customer must observe the safety
recommendations of the supplier of the gas or vapour.
CAUTION
Do not use an EXT Split Flow Pump to pump gases containing more than 20% oxygen. If you do, the
lubricant will polymerise and the pump will fail prematurely.
•Air •Nitrogen •Helium
• Carbon monoxide • Krypton • Hydrogen
•Neon •Argon •Butane
• Ethane • Propane
• Methane • Carbon dioxide