8. Operation
Fig. 58 Rotating specimen stage
1 Object displacement (Y direction)
2 Object displacement (X direction)
3 Torque adjuster (Y-direction)
4 Torque adjuster (X-direction)
5 Focus wheel for fine focusing
8. Operation
8.1 Switching on the unit
When using a gas discharge lamp, the external
supply unit must be turned on separately (56.1).
Switch on the microscope or the CTR5000 at the
power switch.
All motorized microscope components will then
run through an initialization phase.
After the initialization is complete, the display
on the stand (Fig. 57) shows the current micro-
scope setting.
Fig. 57 Display after initialization
8.2 Stages and object displacement
Lengthening the coaxial drive
For lengthening, pull the lower handle (58.2)
downwards. Repeat with the upper handle
(58.1).
Setting the movement rate (torque)
The torque for x and y can be individually adjust-
ed using two knurled rings (58.3, 58.4).
Fig. 56 Front view of the ebq 100 supply unit
1 Power switch
2 Lamp status
1 2
1
4
2
3
5